Transmitter LKM406
The LKM406 is an analog transmitter and can be designed for the Pt100 or Pt1000 measuring resistor according to DIN EN 60751. Its design is ideally suited for mounting in tubular housings. It converts the temperature-dependent resistance value of the sensor into a standard signal of 4...20mA. Lead resistances are largely compensated by means of a three-wire circuit.
The transmitter is supplied calibrated to customer specifications. By means of span and zero controllers, a subsequent fine adjustment can be carried out on site, if not encapsulated.
There are three variants available:
With this arrangement, customer-specific probes with integrated transducer, such as our Type406 or Type409, can be realized.
| Resistance thermometer | Specified range in °C |
|---|---|
| Pt100 | -200 – 850 |
| Pt1000 | -200 – 450 |
| Parameter | Value | Remark |
|---|---|---|
| Input | Pt100 / Pt1000 | 2- / 3-wire configuration |
| Measuring current | approx. 0.8 mA | depending on sensor resistance |
| Zero point | -200 – 500 °C | value for 4 mA |
| Span | 20 – 850 K | zero point + span value 20 mA |
| Linearity error | < 0,1% FS | |
| Loop voltage | 10 – 35 V DC | reverse polarity protected |
| Output | 4 – 20 mA | current loop |
| Sensor break | > 20 mA | |
| Sensor short circuit | < 4 mA | |
| Response time | < 0.1 s | |
| Temperature coefficient | < 100 ppm / °C | |
| Operating temperature range | -40 – 85 °C | |
| Humidity | < 95 % | relative humidity |
| Loop connection | Hirschmann connector type GSP | pins 1 (+) and 2 (-) |
| Sensor connection | exposed solder contacts | |
| Optional potting compound | black polyurethane | |
| PCB dimensions | 26 x 15.5 x 12.5 mm | L x W x H |
| PCB deviation from plumb | 5 ° | |
| Weight | approx. 10 g (unpotted) | |
| EMC | EN 61326-1 EN 61326-2-3 |
emission and immunity |
| Document | Format | Language(s) |
|---|---|---|
| Datasheet – LKM406 Technical Specification |
||
| Operating Instructions – LKM406 Installation and operating manual |
| Program / Package | Format | Language(s) |
|---|---|---|
| Program – AnalogToTemperatur Conversion of analog output values (mA or V) to temperature |
ZIP |
{{.}}
{{/content}}{{{.}}}